Sper-scientific 850055 pH Meter - Advanced Manuel d'utilisateur Page 13

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 40
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 12
13
SETUP MODE
6. Press MN/MX/AV to enter P3.5 and view the offset val-
ue. The offset value is the mV value of pH 7 and the
default offset value is 0.0. The offset value will be
different after calibration. If the value is outside the
range of ± 60 mV, replace the probe.
P4.0 pH Calibration Buffer
This meter allows the selection of two different types of pH
buffers: NIST or CUSTOM. Selection of the proper buffer
type more accurately calibrates the probe to specific
requirements.
NIST buffer: (five settings)
pH 1.68, 4.01, 6.86, 9.18, 12.45
CUSTOM buffer: (five ranges)
pH 1.00-3.00, 3.50-5.50, 6.00-8.00, 8.50-10.50,
11.50-13.50
Select Buffer
1. Press SET for 2 seconds to enter setup.
2. Press to select pH buffer. “BUF” will appear on the
middle of the LCD and P4.0 will appear on the lower
portion.
Vue de la page 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 39 40

Commentaires sur ces manuels

Pas de commentaire